Industrial Turbidity Meter
(0 reviews)
Get Price
(562) Views
(0) Contacts

WGZ series of scattered light turbidity meter is used for measuring suspended in water or in transparent liquids insoluble particulate matter produced by scattering, and quantitative standard to the suspended particle matter content.


Contact Company

Hangzhou Lohand Biological Technology Co.,Ltd
Zhejiang, China
(2215) Views
(18) Contacts

Matters needing attention:
WGZ series of scattered light turbidity meters is a combination of precision measuring instruments, before the operation should carefully read the instructions and through the correct operation can obtain accurate measurement results.
1, the use of the environment must accord with the working conditions.
2, the measuring tank must be a long time clean and dry, without dust, and not be covered with a shading cover.
3, humid climate, must be extended accordingly boot time.
4, the solution to be measured along the sample bottle should be carefully poured, preventing the generation of bubbles, and influencing the measurement accuracy.
5, replacing the specimen bottle or repair shall be calibrated again.
6, a professional repair engineer, does not open the instrument repair.
Working conditions:
1, temperature 5~ 35 ℃
2, relative humidity of not more than 80%
3, the power supply voltage 220V ±22V 50± 0.5Hz frequency, and has the good grounding.
4, the instrument should be placed horizontally on a smooth test bench, effectively avoiding direct light
 5, around the tool, should have sufficient space to facilitate heat dissipation, and no strong vibration source and strong magnetic field interference Around 
6, the air should be no dust or corrosive gas.
Overhaul:
Failure phenomenon
Possible causes
Repair method
1, after starting up, display
A、power line and socket connection is bad or loose
A、fastening socket or for a power line
 
2, measuring reaction
A light damage
B, internal connector lose
C, electrical system fault
A, replacement
B, fastening
C, maintenance
 
 
 
3, measurement instability or drift
A, a solution with bubbles or particles in does not stop moving
B, circuitry within the apparatus of damp
C, the outer surface of the bottled water sample
D, external interference
E, the supply voltage instability of
A re-sampling or extended reading time
B, prolonging start preheat time preheating dehumidification
C, dry specimen bottle
D, eliminating the interference source
E, exclusion of unstable factors
 
 
4. zero, less than zero
A, adjustable zero without the use of zero turbidity water
B, zero potentiometer damage
C, zero range migration
D, light path offset
A should adopt zero turbidity water
B, replacement potentiometer
C, W2 potentiometer adjustment board
D, adjustment
 
5, not to the correction value
A, standard solution standard value is not accurate
B, correction potentiometer is damaged
C, range migration correction
D, light path offset
 
A. accurate preparation of standard solution
B, replacement
C, W1 potentiometer adjustment board
D, adjustment
 
6. short stroke
A, due to the transport screen vibration caused by poor contact
B, LCD LCD screen damage
C, circuit board fault
A, check
B, replacement
C, maintenance


Product name turbidity measurement device
Model WGZ-20
Characters Bench top type
Minimum Readout ( NTU) 0.1
Power Supply AC220V/50Hz
Measuring Range(NTU) 0--200
Basic Error(F.S) ±2.5%
Zero Draft(F.S ) ±0.2%
Repeatability ≤1.5%

Can be widely used in power plants, pure water plants, water plant sewage treatment factories, beverage factories, environmental protection departments, industrial water, wine making industry and the pharmaceutical industry, hospitals, epidemic prevention departments, and other departments of the determination of turbidity.