  • This halogen acid gas content test is used to test the acidity of the gas that is released during combustion from cable or optical cable components material; The equipment is designed and produced on the basis of IEC 60754.
  • The cable material is processed by a Prius tube furnace, and then the pH value and conductivity of the gas that is generated in the combustion process are measured, finally, the cable material of acidic, as well as the concentration of halogen acid gas volume, PH, conductivity, etc, can be determined.
Standards:
• IEC 60754-1 Ed. 2.0 b:2011:specifies the apparatus and procedure for the determination of the amount of halogen acid gas, other than hydrofluoric acid, evolved during the combustion of compounds based on halogenated polymers and compounds containing halogenated additives taken from electric or optical fiber cable constructions
• IEC 60754-2 Ed. 1.0 b:2011:specifies the apparatus and procedure for the determination of the potential corrosivity of gases evolved during the combustion of materials taken from electric or optical fiber cable constructionsby measuring the acidity (pH) and conductivity of an aqueous solution resulting from the gases evolved during the combustion.


Place of Origin CHINA
Model Number CTA
Brand Name AI
Certification ISO
Test Standard IEC 60754-1&2
Power Supply AC 220V, 50 / 60Hz, 15A
Gas Air And Propane
Size 1,500 (W) X 500 (D) X 600 (H) Mm
Weight 60kg
Keyword Halogen Acid Gas Content Test Apparatus

1, The Halogen Acid Gas Content Test Apparatus is designed with an optimized structure, and logical layout, and is easy to operate.
2, Al the fixtures are made of stainless steel.
3, The effective length of the tube-type furnace is 500~600mm, inner diameter is 40~60mm, with adjustable electric systems.
4, Halogen Acid Gas Content Test Apparatus is equipped with an imported thermocouple from the United States.
5, The size of the combustion boat is a length of 45~100mm, a width of 12~30mm, and a depth of 5~10mm.
6, The electrical conductivity measuring range is 0.01uS/mm~100uS/mm.
7, The control system uses electronic temperature control and intelligent control tube furnace temperature.
8, The test apparatus is equipped with two gas collection bottles.
9, The size of the sample chamber is 1 liter.
10, Halogen Acid Gas Content Test Apparatus is supplied with a PH value and conductivity detector, with an electronic display and an electrode.
11, the Air flow meter is equipped with a silicon dioxide fire pipeline to prevent corrosion by the gas.
12, Activated carbon (air filter): filter air through a filter.
13, silicone is used for drying air sources.
14, Pyrex pipeline: Users can easily insert samples into the Prius tube to protect the pipeline from damage.
15, Combustion boat moving system: combustion boat moves automatically to achieve a more intelligent test.
16, The mode switching: mode switch (two test methods to choose from, IEC 60754-1 & IEC 60754-2).
17, The temperature alarm system: control the furnace temperature against overheating.
18, Timer: IEC 60754-1 or IEC 60754-2 time measurement device.
19, Drawer: used to store parts of the box (such as conductivity meters, pH meter, etc.)
20, Temperature indicator: shows the furnace core temperature.
21, Permanent magnetic stirrer: The agitator is used to stir the combustion gases in distilled water.
22, Conductivity meter: An instrument used to measure the combustion gas conductivity.